This universal design allows the handle to accept any standard Welch Allyn blade. New in open box. Welch Allyn 2.5V Power Handle for Standard Laryngoscopes - Model 60200 eBay. Welch Allyn standard laryngoscope handles. Maintenance instructions. Replace the lamp. WARNING: Lamps, if left illuminated, could generate sufficient heat to cause burns. CAUTION: Use only Welch Allyn replacement lamps REF: 06000 to ensure proper illumination alignment. Unscrew bottom cap of handle counterclockwise and remove batteries per. Fiber Optic Laryngoscope Handles HPX, User Manual Cross Reference Other Manufacturers Fiber Optic Laryngoscope Handles/Blades, Reference Fiber Optic and Standard Laryngoscope Handles, Troubleshooting Article. Part #: 69097 GTIN: 1283. Welch Allyn English MacIntosh 2.5 V Standard Laryngoscope Set; Lamp at Distal Tip; Stainless Steel, Blade Sizes #1 to #4 (Qty. 4), Medium C-Cell Power Handle, Small AA-Cell Power Handle, Hard Storage Case. This universal design allows the handle to accept any standard Welch Allyn blade. New in open box. Welch Allyn 2.5V Power Handle for Standard Laryngoscopes - Model 60200 eBay.
Advertising
Directions for use
Welch Allyn fiber optic laryngoscope handles
3
Cleaning and
intermediate level
disinfection:
intermediate level
disinfection:
Drying:
Maintenance,
Inspection and Testing
1. Follow the germicidal wipe manufacturer’s instructions
to clean all exposed surfaces of the main handle, end
cap, and lamp cartridge.
cap, and lamp cartridge.
2. If necessary, brush with a dry, soft-bristled brush and
re-wipe to loosen/remove excessive visible soil.
3. After all visible soil is removed, re-wipe to wet all
surfaces and allow adequate contact time for
disinfection as directed by the germicidal wipe
manufacturer.
disinfection as directed by the germicidal wipe
manufacturer.
CAUTION: Only use quaternary ammonium
isopropanol based germicidal wipes.
Allow components to air dry.
1. Inspect each component area (per figure 3) for damage or
deterioration.
WARNING: Discard any component that shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble handle (per figure 4) with new or
batteries in known good condition.
3 . Attach handle to a clean and disinfected test blade
assembly in known working condition. Verify that:
•
Blade assembly engages and locks onto handle.
•
Blade assembly deploys into its locked position on
handle AND lamp illuminates.
handle AND lamp illuminates.
•
Light output is satisfactory.
If the lamp fails to light or output is low, check or
replace the lamp and/or batteries.
replace the lamp and/or batteries.
Storage
Store handle per facility practice to allow device to remain
clean, dry, and ready for service.
clean, dry, and ready for service.
End of reprocessing instructions for intermediate level disinfection.
Advertising
Popular Brands
Popular manuals
Advertising
6
Directions for use
Welch Allyn standard laryngoscope handles
Maintenance instructions
Replace the lamp
WARNING: Lamps, if left illuminated, could generate sufficient heat to
cause burns.
cause burns.
CAUTION: Use only Welch Allyn replacement lamps REF: 06000 to ensure
proper illumination alignment.
proper illumination alignment.
1. Unscrew bottom cap of handle counterclockwise and remove batteries per figure 3.
2. Remove lamp cartridge assembly from main handle by applying finger pressure in the
2. Remove lamp cartridge assembly from main handle by applying finger pressure in the
direction shown by the arrow per figure 4.
3. Remove shroud cap from shroud by rotating shroud cap counterclockwise per figures 5
and 6.
4. Rotate lamp counterclockwise to remove per figure 6.
5. Replace lamp and rotate shroud cap clockwise to tighten per figure 7.
6. To replace lamp cartridge assembly in the main handle, invert handle, then gently slide
5. Replace lamp and rotate shroud cap clockwise to tighten per figure 7.
6. To replace lamp cartridge assembly in the main handle, invert handle, then gently slide
Standard Laryngoscope Handles User Manual Welch Allyn Welch Allyn Manual
the cartridge down the inside of handle, tipping it side to side until the shroud cap
exits opening on top.
exits opening on top.
7. Re-insert batteries and apply slight pressure to set the cartridge in place.
8. Replace and tighten bottom cap.
9. Reprocess repaired assembly as appropriate per these instructions.
8. Replace and tighten bottom cap.
9. Reprocess repaired assembly as appropriate per these instructions.
Figure 3 Metro c5 3 series user manual.
Figure 4
Figure
5
Figure
6
1. Shroud cap
2. Lamp
3. Shroud
2. Lamp
3. Shroud
Figure 7